Medical/Commercial Transient Generators

Meet conducted immunity requirements in IEC/EN standards such as IEC 60601 and EN 550022. The EMC Shop is an ISO 17025 accredited company. Contact us to rent of buy Medical/Commercial Transient Generators, most equipment is in stock and can ship same day.

Medical/Commercial Transient Generators There are 3 products.

Subcategories

  • Combination Wave/Surge Generators

     

    Combination wave generators are used to meet IEC/EN 61000-4-5ANSI (IEEE) C62.41 and other surge immunity standards. The EMC Shop has a wide selection of combination wave generators and associated coupling/decouping networks in stock.

     

    Also See - IEC 61000-4-5 Surge Immunity Testing

    Also See - IEC 61000-4-4 EFT Testing

    Also See - IEC 61000-4-6 Testing and Setup

    Video Overview - Surge4-5 Combination Wave Generator

    Getting Started witha Combination Wave Generator

    Combination wave/surge generators are typically benchtop, but heavy equipment. Testing on data and three phase power lines require additional coupling/decoupling networks. Pre programmed test sequences for surge immunity allows you to begin testing once coupling in your EUT and properly grounding the test setup.

  • Electrical Fast Transient EFT/Burst Generators

     

    The EMC Shop EFT/Burst equipment for compliance and pre-compliance testing. Rent or buy stand alone electrical fast transient generators or try a multifunction generator to add SurgePower Fail and other electrical disturbance tests.

    Get Application Support, Pricing, Lead Time and More: Contact Form

     

    Electrical Fast Transient (EFT)/Burst generators are required for conducted immunity testing standards, such as IEC/EN 61000-4-4.  Electrical Fast Transient EFT/Burst events are small electrical surges caused by a local source, typically an electrical switch, it can occur within a product itself as well as from the data or power lines in the facility.

     

    Electrical Fast Transient EFT/Burst Generator Setup

     

    Also see - EFT Testing

  • Surge & Burst/EFT Generators

    Conducted immunity testing is a major requirement to attain CE marking for products. Multiple types of electrical disturbances are required to be tested and certified to pass the internal compliance standards, EFT/Burst standard EN/IEC 61000-4-4 and Surge (combo wave) standard IEC/EN 61000-4-5 are available in most standard EMI test sets made today. The EMC Shop stocks all types of Surge & Burst/EFT Generators for compliance and pre-compliance testing. Rent or buySurge & Burst/EFT Generators for individual tests or multifunction generators to complete an assortment of tests in a compact solution. Most systems can be easily upgraded to include additional tests as required.

    Get Application Support, Pricing, Lead Time and More: Contact Form



    Transient generators are typically benchtop, but heavy equipment. When doing testing on three phase power lines systems usually require much larger coupling/decoupling networks which substantially increase system size. EM test environments with ground planes with connections to earth are required to safely use transient generators.

    Common Transient Tests

    Combination Wave/Surge:

    • IEC/EN 61000-4-5 specifies two Combination Wave pulses: Open Circuit Voltage Short Circuit Current
    • A common combination wave test is the 1.2/50 & 8/20
    • Another common wave test is the 10/700 & 5/320, or otherwise known as telecom wave
    • The four standard test levels , excluded of EUT: 0.5, 1, 2, 4 kV (0.25, 0.5, 1, 2 kA)

    Burst/EFT:

    • A typical burst has a single pulse rise time/duration of 5 ns/50 ns from a source impedance of 50 Ω
    • Bursts of 15 ms duration with a repetition rate of 5 kHz (or 100 kHz) are applied every 300 ms
    • The standard defines four basic test levels, all with repetition frequencies 5 or 100 kHz
    • The level voltage depends on equipment being tested:
    • Power ports: 0.5 , 1, 2, 4 kV
    • Signal and control ports: 0.25, 0.5, 1, 2 kV

  • Surge, Burst/EFT, Ring Wave, Dips & Interrupts

    Accomplish all common conducted immunity testing to conform with EMC (electromagnetic compliance) standards for CE marking in one multi-purpose instrument. Generate electrical disturbances for Surge (Combination wave pulse 1.2/50 - 8/20 μs), Burst/EFT (5/50 ns), and Ringwave (0.5 μs/100 kHz) as well as pass power quality compliance standards by simulating AC dips, drops, and interruptions. The EMC Shop stocks all types of multifunction generators for compliance and pre-compliance testing. Rent or buy generators for individual tests or multifunction generators to complete an assortment of tests in a compact solution. Most systems can be easily upgraded to include additional tests as required.

    Get Application Support, Pricing, Lead Time and More: CONTACT FORM



    Multifunction generators are typically benchtop, but heavy equipment. When doing testing on three phase power lines systems usually require much larger coupling/decoupling networks which substantially increase system size. EM test environments with ground planes with connections to earth are required to safely use transient generators.



    Common Combination Wave/Surge Tests

    • IEC/EN 61000-4-5 specifies two Combination Wave pulses: Open Circuit Voltage Short Circuit Current
    • A common combination wave test is the 1.2/50 & 8/20
    • Another common wave test is the 10/700 & 5/320, or otherwise known as telecom wave
    • The four standard test levels , excluded of EUT: 0.5, 1, 2, 4 kV (0.25, 0.5, 1, 2 kA)

    Common (EFT)/Burst Generator Tests

    • A typical burst has a single pulse rise time/duration of 5 ns/50 ns from a source impedance of 50 Ω
    • Bursts of 15 ms duration with a repetition rate of 5 kHz (or 100 kHz) are applied every 300 ms
    • The standard defines four basic test levels, all with repetition frequencies 5 or 100 kHz
    • The level voltage depends on equipment being tested:
    • Power ports: 0.5 , 1, 2, 4 kV
    • Signal and control ports: 0.25, 0.5, 1, 2 kV

    Ring Wave Immunity Generator Tests

    For compliance, the ring wave has a measured 0.5 us rise time and 100 kHz oscillation frequency This impulse is commonly used for applications related to power distribution and railway environments.

  • Magnetic Field Generators

    Conducted immunity testing is a major requirement to attain CE marking for products. Magnetic pulse field strengths conforming to EN/IEC 61000-4-8 and EN/IEC 61000-4-9 are the most common EMI magnetic test. The EMC Shop stocks magnetic field generation equipment for compliance and pre-compliance testing. Rent or buy magnetic field generators for individual tests or multifunction generators to complete an assortment of tests in a compact solution. Most systems can be easily upgraded to include additional tests as required.

    Get Application Support, Pricing, Lead Time and More: CONTACT FORM

    Common Magnetic Field Generation Tests

    • magnetic disturbances at power frequencies 50 Hz and 60 Hz related to
    • residential and commercial locations
    • industrial installations and power plants
    • medium voltage and high voltage sub-stations

     

  • Flicker & Harmonics Generators

    Flicker and harmonics generators are required to perform compliance testing to IEC 61000-3-2, 3-3, 3-11, and 3-12. The EMC Shop has power sources available for rent or sell to perform this testing. The EMC Shop stocks all types of Flicker and harmonics generators for compliance and pre-compliance testing. Rent or buy Flicker and harmonics generators for individual tests or multifunction generators to complete an assortment of tests in a compact solution. Most systems can be easily upgraded to include additional tests as required.

    Get Application Support, Pricing, Lead Time and More: CONTACT FORM



    Flicker and harmonics generators are typically benchtop, but heavy equipment. When doing testing on three phase power lines systems usually require much larger coupling/decoupling networks which substantially increase system size. EM test environments with ground planes with connections to earth are required to safely use transient generators.



    Common Flicker and Harmonics Generator Tests

     

Showing 1 - 3 of 3 items
    • 4-Quadrant – Source and sink current
    • Up to 25 Amp DUT
    • Perform Dropout, Surge, Ripple Simulations
    • Large Standards Library
    • Automotive, Military, Aerospace and Commercial
    In Stock
    Request Quote/More Info
    • Multiple configurations in stock
    • Rentals ship same day
    • ISO 17025 calibration available
    • 3 Phase Coupling and Data Line Clamps Available
    In Stock
    Request Quote/More Info
    • 4-Quadrant – Can source and sink current
    • ±80V – DC supply for 12V to 48V systems; meets 80V surge requirements
    • 300 kHz Sine – DC ripple tests for all major standards
    • 3µS Rise Time – Exceeds surge and dropout slew rate requirements
    • 3mΩ DC Source Impedance – Out-performs ISO 7637-2 requirements