CS116 Purpose
Conducted susceptibility, damped sinusoidal transients, cables and power leads.
This requirement is applicable from 10 kHz to 100 MHz for all interconnecting cables, including power cables, and individual high side power leads. Power returns and neutrals need not be tested individually.
This test procedure is used to verify the ability of the EUT to withstand damped sinusoidal transients coupled onto EUT associated cables and power leads.
Test Equipment
The test equipment shall be as follows:
a. Damped sinusoid transient generator, ≤ 100 ohm output impedance
b. Current injection probe
c. Oscilloscope, 50 ohm input impedance
d. Calibration fixture: Coaxial transmission line with 50 ohm characteristic impedance, coaxial connections on both ends, and space for an injection probe around the center conductor
e. Current probes
f. Waveform recording device
g. Attenuators, 50 ohm
h. Measurement receivers
i. Coaxial loads, 50 ohm
j. LISNs