Viewed products
Full-vector network and spectrum...
M1, M2 Complex Signal Analyzer...
Refurbished
Warning: Last items in stock!
Availability date:
Agilent / HP 4395A provides excellent vector network, spectrum, and optional impedance measurements for lab and production applications. Gain, phase, group delay, distortion, spurious, CN, and noise measurements often required for evaluating components and circuits can be measured using one instrument.
When combined with a test set, the 4395A provides reflection measurements, such as return loss, SWR, and S parameters. As a vector network analyzer, the 4395A operates from 10Hz to 500MHz with 1 mHz resolution; and its integrated synthesized source provides -50 to + 15dBm of output power with 0.1 dB resolution. The dynamic magnitude and phase accuracy are +/-0.05dB and +/-0.3deg. so that it can accurately measure gain and group delay flatness, which are becoming more important in modern electronics systems.
As a spectrum analyzer, the 4395A operates from 10 Hz to 500 MHz with resolution bandwidths (RBWs) spanning 1 Hz to 1 MHz in a 1-3-10 steps. A fully-synthesized local oscillator allows stable and accurate frequency analysis. Direct A/D conversion (no LOG amplifier is used) results in +/-0.8 dB level accuracy (@50 MHz, -20 dBm). Noise sidebands fall below -100 dBc/Hz @ 100 kHz offset from carriers, while sensitivity is -145 dBm/Hz at 10 MHz.
New Approach to Electronic Component and Circuit Evaluation
A new concept in testing that combines the power of a vector network analyzer, a spectrum analyzer, and an optional impedance analyzer.
Take an innovative approach to evaluating electronic components and circuits. The Agilent Technologies combination analyzers family combines three analyzer functions in one powerful instrument: a vector network analyzer, a spectrum analyzer, and an optional impedance analyzer. Now you can use one analyzer for multiple testing needs.
Because many electronic components such as amplifiers require both vector network and spectrum parameters, you need a combination analyzer to meet new testing requirements such as:
Compared with using separate analyzers, you save equipment cost and bench space with the Agilent 4395A/96B, and avoid time-consuming and messy signal cabling to multiple instruments.
SPECIFICATIONS | |
Network Analyzer | |
Frequency Range | 10 Hz to 500 MHz1 |
Frequency Resolution | 1 mHz |
Output Power range | –50 to 15 dBm |
Dynamic Range | 115 dB @ 10 Hz IFBW |
Dynamic Accuracy | ±0.05 dB/0.3 deg. |
Calibration | Full two-port |
Spectrum Analyzer | |
Frequency Range | 10 Hz to 500 MHz |
Noise Sidebands | < –104 dBc/Hz typical at 10 kHz offset |
Resolution Bandwidth | 1 Hz to 1 MHz in 1-3-10 steps |
W Dynamic Range | > 100 dB third-order free dynamic range |
Level Accuracy | ±0.8 dB @ 50 MHz |
Sensitivity | –145 dBm/Hz @ freq. = 10 MHz |
Impedance Analyzer 2 | |
Frequency Range | 100 kHz to 500 MHz |
Meas. Parameter | |Z|, θz, R, X, |Y|, θy, G, B, Cs, Cp, Ls, Lp, Rp, Rs, X, D, Q, IΓI, Γx ,Γy |
Z Accuracy | ±3% (typical, basic accuracy) |
Source Level | –56 dBm to +9 dBm (at DUT) |
DC bias | ±40 V (20 mA (max)) (Option 4395A-001 DC source or external DC source is required.) |
Compensation | OPEN/SHORT/LOAD port extension |
Standard features: Instrument BASIC, GPIB port, 3.5" floppy disk drive, direct print, RAM disk, VGA monitor output. Optional features: Impedance measurement (Option 4395A-010), time-gated spectrum analysis (Option 4395A-1D6), high-stability frequency reference (Option 4395A-1D5), 50 Ω to 75 Ω spectrum input impedance conversion (Option 4395A-1D7), DC source (±40 V, 100 mA (ALC)) (Option 4395A-001). | |
1. 100 kHz to 500 MHz if using the Agilent 87511A/B S-parameter test set. 2. With Option 4395A-010 and the Agilent 43961A RF impedance test kit |