Voltage Rating | 500 Volts rms (line to line), 288 Volts rms (line to ground) Max DC Voltage: 705V |
Frequency Range | 9 kHz to 30 MHz |
RF Connector | RF Measurement Port 50Ω - N-Type (female) |
Standards Met | ANSI C63.4, CISPR 16-1-2 |
Inductors | 50 µH |
Mains & EUT Connections | Remote Interface Power Input Port 5.5/2.5 mm Power Jack |
Current Rating | 16 Amperes |
Insertion loss | less than 0.7 dB |
Application/Test Type | Power Line Conducted Emissions Tests |
Isolation | more than 40 dB |
Dimensions | 19.3” x 13.9” x 2.3" |
Weight | 27.5 lbs. |
The Com-Power LI-3P-216 LISN provides the necessary measurement platform for performing power line conducted emissions compliance testing as required by most worldwide standards for commercial products. The LISN will perform each of the following functions during the measurement:
The LI-3P-216 LISN uses air-core inductors to prevent saturation and permeability variation. The mounting plates are left unpainted in order to facilitate connection to earth ground in their installation, which is essential due to high leakage currents.
The side panels for the LISN are louvered for cooling purposes.
Remote or Local Operation:
Remote switching of the line under test (L1, L2, L3, N) is performed using the RLI-100 Remote LISN Interface, which controls the LISN via fiber optic connection.
In addition to the remote method, the line under test can also be selected using the mechanical, four-position switch located on the front panel of the LISNs.
Using either switching method, the lines which are not selected are internally terminated into 50 ohms, while the selected line is terminated by the 50 ohm input impedance of the measuring instrument.
Transient Protection:
The Com-Power LIT-153A Transient Limiter (see accessories below) is a recommended accessory for protection of the RF input of your measuring instrument from potentially damaging, instantaneous voltage transients.
The transient limiter also reduces the possibility of overload by incorporating two 5 dB attenuation/impedance matching pads, in addition to its low-pass and high-pass filter sections which further attenuate any out-of-band emissions.
Calibration:
Each LISN is individually calibrated in compliance with the relevant requirements of CISPR 16-1-2 and ANSI C63.4. Impedance, Phase, Isolation, and Insertion Loss data is supplied with each unit, along with the certificate of calibration.
Recognized ISO 17025 accredited calibration is also available upon request.