The ESD Simulator SESD 230 is suitable for performing EMC tests on systems in accordance with the standard IEC / EN 61000-4-2 (ESD test). Higher test levels can be set far beyond the standard limits. Depending on the test object and test setup, two test methods are to be used:
The contact discharge is the favourite test method since it is most reproducible. Air discharges are used when contact discharges are not possible - e.g. at plastic housings. The test voltages defined for each test method are shown in the table below:
|
Schloder SESD 230 Technical Data | |
Generator | |
Output voltage, adjustment via digital potentiometer | |
Test mode air discharge | 0,2 kV to 30 kV, 100V steps |
Test mode contact discharge | 0,2 kV to 30 kV, 100V steps |
Polarity of the output voltage | Positive and Negative |
Test modes | Air- and Contact Discharge |
Repetition frequency of the discharge pulses | |
Air discharge | Single pulse or repeated * *(frequency depends on the distance between the discharge electrodes and the examinant) |
Contact discharge | Single pulse, 0,1 Hz, 0,2 Hz, 1 Hz, 2 Hz, 5 Hz, 10 Hz, 20 Hz |
Continuous operation | Possible at air- and contact discharge |
Holding time | ≥ 5 sec |
Pre selectable counter | 1 - 9999 |
Discharge electrodes | In conformity to IEC / EN 61000-4-2 |
Energy storage capacity | 150 pF ± 10% |
Discharge resistor | 330 Ohm ± 5% |
Operation temperature range | 0 - 40° Celsius |
Relative humidity | 0 - 60% |
Weight | App. 1470 g |
Power supply | |
Supply voltage | IN: 100-240 VAC / 47-63 Hz; OUT: 9 VAC / 3 A Weight: app 200 g |