Fully IEC 61000-4-5 compliant...
Used
Typically In Stock
Warning: Last items in stock!
Availability date:
Supply Voltage | 115/230V +10/-15% |
Mains & EUT Connections | External Coupler Only |
Pulse voltage (open circuit) | 500V - 10,000V ± 10% |
Combination wave pulse 1, 2/50 - 8/20 μs (Hybrid-Surge pulse) Pulse conforms to IEC/EN 61000-4-5 | |
Parameter | Value |
Pulse current (short circuit) | max. 5,000A ± 10% |
Pulse repetition | 1s - 999s, depending on the voltage |
Polarity | pos / neg / alternate |
Phase synchronization | asynchronous, synchronous 0 to 359º (in 1º steps) |
Test duration | 1 - 30,000 or endless |
Dimensions | 19"/6HU; |
Weight | approx. 35kg |
Surge pulses occur due to direct or indirect lightning strokes to an external (outdoor) circuit. This leads to currents or electromagnetic fields causing high voltage or current transients. Another source for surge pulses are switching transients originating from switching disturbances and systems faults. Due to the characteristic of the phenomenon nearly every electrical and electronical device may suffer from such lightning events which justifies the necessity of surge tests being widely performed. Surge voltage can reach several thousands of volts and surge current is seen to reach several thousands of amps.