M Series

Quick Links

Information

Viewed products

Com-Power CDN-M225E Coupling Decoupling Network for Unscreened Power Supply Lines

New

  • 150 MHz - 230 MHz Fequency Range
  • Designed for IEC / EN 61000-4-6
  • Two conductor power cable
  • Individual Calibration Included
  • Three Year Warranty

More details

Typically In Stock

Request Quote/More Info

PDF Downloads

Specifications

Voltage Rating250 V AC / 350 V DC (Line to Ground)
Frequency Range150 kHz to 230 MHz
Standards MetIEC / EN 61000-4-6
RF Connector50 Ω BNC (Female)
Mains & EUT Connections4 mm shrouded banana sockets
Common Mode Impedance150 kHz - 26 MHz: 150Ω ± 20Ω 26 MHz - 80 MHz: 150Ω + 60Ω / – 45Ω 80 MHz - 230 MHz: 150Ω + 60Ω / – 60Ω
Current Rating25 Amps AC / 17 Amps DC (Max)
Application/Test Typetwo conductor power cable
RF Input Voltage40 V (Max)
Voltage Division Factor9.5 dB +4 / -1
Decoupling of Common Mode Disturbance≥ 40dB (EUT/AE)
Dimensions6 x 6 x 13 inches 15.2 x 15.2 x 33 cm
Weight5 lbs. 2.3 kg

Test Equipment Description

The CDN-M225E is designed for testing products that uses two conductor power cable. It has two 4 mm shrouded banana socket for both EUT and AE power connections. The CDN-M225E can handle up to 25 Amps AC and 17 Amps DC of current. The RF disturbance signal coupling port is female BNC. It can handle up to 40V of RF input Voltage. The bottom surface of the CDN is not painted for easy and effective grounding.

All Com-Power CDNs are individually calibrated. The Com-Power CDN-M225E fully complies with the requirement contained in the IEC 61000-4-6 and CISPR 16-1-2. All Com-Power CDNs can be purchased separately or as part of the CIS series conducted immunity test system. This is a pre-packaged solution that includes an ACS series power amplifier and all accessories required for the test.

Application

During conducted Immunity testing, CDNs are utilized to provide a means of coupling RF common mode signals to each line. In addition, CDNs provide the required common mode impedance to the EUT, isolation to the auxiliary equipment via common mode decoupling of the disturbance signals and provide uninterrupted communication between the EUT and auxiliary equipment.

Before you begin testing with the CDN-M225E you will need to establish calibrated drive levels corresponding to your desired test levels. During drive level calibration, the RF signal level being injected to the CDN is adjusted incrementally until the voltage level measured at the 150Ω to 50Ω adapter (ADA-515-2) connected to the EUT port is approximately equal to the Umr value given in the table below. The ADA-515-2 and accessories needed for this test are available from Com-Power.

Test Levels
Open Circuit Voltage
Umr
10.167
30.5
101.67

Umr= Voltage level measured at the output of the 150Ω to 50Ω adapter (ADA-515-2)

Typical Data

CDN M-225E Accessory Kit:

Associated EMC Test Equipment

9 other products in the same category: