Magnetic H-Field Probes

Quick Links

Information

Viewed products

Hioki Magnetic Field Field HiTester 10 Hz to 400 kHz View larger

Hioki Magnetic Field Field HiTester 10 Hz to 400 kHz

New

  • Complies with ICNIRP 2010 guidelines as well as other relevant standards for evaluation testing.
  • Complies with IEC 62110/IEEE 644 as well as IEC 62233.
  • Bundled with 3 cm2 Sensor used for magnetic field distribution analysis, and 100 cm2 Sensor used with the IEC/EN 62233 standard analysis
  • User-selectable display units (T, A/m, and G)
  • Simple operation for easy measurement
  • Bundled with PC application software
  • Level output for RMS value, or 3-axis waveform output for magnetic fields

More details

Request Quote/More Info

PDF Downloads

Test Equipment Description

Hioki magnetic field meters are designed to measure magnetic flux density and level of magnetic field exposure, and can be used to assess conformance to ICNIRP 2010 and IEC/EN62233 and in research on magnetic field exposure. The FT3470-52 measures magnetic flux density and magnetic field exposure generated by products manufactured at a customer site or produced by devices in a customer environment, and comes bundled with 2 sensors.

Magnetic flux density (Bandwidth)10 Hz to 400 kHz/ 10 Hz to 2 kHz/ 2 kHz to 400 kHz
Exposure levelGeneral Public/ Occupational
DisplaySingle axes X, Y, Z (2000 counts), Composite RMS value R (3464 counts), Magnetic flux density (unit: T, G, A/m), Exposure level (unit: %)
Magnetic flux density/Ranges and accuracy[X, Y, Z axes] Effective measuring ranges: 2.000 μT to 2.000 mT, 4 ranges, Accuracy: ±3.5% rdg. ±0.5% f.s.
[R axis] Effective measuring ranges: 3.464 μT to 3.464 mT, 4 ranges, Accuracy: ±3.5% rdg. ±0.5% f.s.
[Valid measurement frequency range]
at 10 Hz-400 kHz mode: 50 Hz to 100 kHz
at 10 Hz-2 kHz mode: 50 Hz to 1 kHz
at 2 kHz-400 kHz mode: 5 kHz to 100 kHz
Exposure level/Ranges and accuracy[X, Y, Z axes] Effective measuring ranges: 20.00 % to 200.0 %, 2 ranges
[R axis] Effective measuring ranges: 34.64 % to 346.4 %, 2 ranges,
Accuracy: Smoothed edges 50 Hz to 1 kHz ±3.5% rdg. ±0.5% f.s.
Accuracy: Smoothed edges 1 kHz to 100 kHz ±5.0% rdg. ±0.5% f.s.
Interfaces[Supporting output] Resultant RMS level output, Exposure level output, Waveform output of magnetic flux density X/ Y/ Z each axis, Output rate: 0.1 mV/display value count
[USB 1.1] Data saving with the PC application
Other functionsMemory function: Up to 99 measured value data, Slow function, Holds the maximum value, Auto power off, Buzzer sound on/off
Power supplyLR6 (AA) alkaline battery ×4, 0.8 VA (at battery operation), Continuous use of 10 hr, or AC adapter 9445-02 (1.0 VA max. consumption)
Dimensions and massMain unit: 100 mm (3.94 in) W × 150 mm (5.91 in) H × 42 mm (1.65 in) D, 830 g (29.3 oz), (including batteries)
100 cm2 Sensor: φ122 mm (4.80 in) × 295 mm (11.61 in)L, 220 g (7.8 oz)
3 cm2 Sensor: □ 27 mm (1.06 in) × 165 mm (6.50 in)L, 95g (3.4 oz)

Associated EMC Test Equipment