AF-Series

Quick Links

Information

Viewed products

Schwarzbeck CDN AF2 Coupling Decoupling Network View larger

Schwarzbeck CDN AF2 Coupling Decoupling Network

New

  • 150 kHz - 230 MHz
  • 4 mm Safety Lab Connectors
  • 250 V AC / 250 V DC
  • 1.5 A
  • CDN for 2 Unscreened Unbalanced Lines

More details

Request Quote/More Info

PDF Downloads

Test Equipment Description

The CDN AF2 is part of the coupling decoupling network series for conducted common mode immunity testing according to IEC 61000-4-6. The CDN AF2 is compliant to IEC 61000-4-6 Annex D3 and is suitable for immunity testing on unscreened and unbalanced control lines with moderate currents.

The CDN AF2 is intended to inject common mode disturbance signals to control lines in the frequency range 150 kHz to 230 MHz. The circuitry is shown in Fig. 5.

All CDNs comply to the requirements of IEC 61000-4-6. Each CDN comes with individually measured data and a calibration certificate for the common mode impedance and the voltage division factor. Further typical data can be found in fig. 3 and fig. 4.

A variety of calibration adapters and other accessories is available as option.

Important Note: The optional shorting adapters must not be used if the CDN is connected to mains! The capacitors at the AE-side are not equipped with discharging resistors

The connection to ground can be accomplished using the ground plane of the CDN. Additionally there is a M4 thread located at the AE side to ground as well as a 4 mm socket to connect the device to ground.

To improve the operational safety the connections are carried out as security sockets. We recommend to use special 4 mm security plugs if you plan to design your own adapters. Those connectors can be purchased optionally.

Fig. 1: typ. EuT common mode impedance
Fig. 2: typ. phase EuT-Port, common- mode
Fig. 1: typ. EuT common mode impedanceFig. 2: typ. phase EuT-Port, common- mode
Fig. 3: typical voltage division factor RF-port to EuT port
Fig. 4: typ. isolation AE to RF-port
Fig. 3: typical voltage division factor RF-port to EuT portFig. 4: typ. isolation AE to RF-port

Schwarzbeck CDN AF2 Principal Circuitry
Fig. 5: Principal Circuitry CDN AF2

Specifications
Frequency range0.15 … 230 MHz
Common mode impedance150 kHz – 24 MHz: 150 Ω ±20 Ω
24 MHz–80 MHz: 150Ω +60Ω / -45 Ω
80 MHz – 230 MHz: 150 Ω ±60
Max. RF-test voltage (emf)30 V
RF-input-power6 W (continuous)
RF-input connector50 Ω BNC (female)
Voltage division factor RF-input – EuT-port9.5 dB 150 kHz – 80 MHz: ±1dB 80 MHz – 300 MHz: +3 dB / - 2 dB
Input Voltage EuT (AE)100 VAC / 150 VDC
Current rating4 A, DC
Housing materialAluminium
Housing dimensions216 x 105 x 108 mm
Weightca. 1320 g
EuT, AE connectors4 mm safety banana jacks
Included accessoriesCA 2/4
3 Pin shorting bar(2 units included by default)
Optional accessories
30 mm distance adapter
50 Ω to 150 Ω adapter

SR 30/4
SR 100-6W

Associated EMC Test Equipment

16 other products in the same category: