Schwarzbeck CDN AF6 Coupling Decoupling Network View larger

Schwarzbeck CDN AF6 Coupling Decoupling Network

New

  • 150 kHz - 80 MHz
  • 4 mm Safety Lab Connectors
  • 63 V AC / 100 V DC
  • 1.5 A
  • CDN for 6 Unscreened Unbalanced Lines

More details

Request Quote/More Info

PDF Downloads

Test Equipment Description

The CDN AF6 is part of the coupling decoupling network series for conducted common mode immunity testing according to IEC 61000-4-6. The CDN AF6 is compliant to IEC 61000-4-6 (comparable to figure D.6) and is suitable for immunity testing on unscreened unbalanced lines.

The CDN AF6 is intended to inject common mode disturbance signals to EuTs with unscreened unbalanced lines in the frequency range from 150 kHz to 230 MHz. The circuitry is shown in Fig 5.

All CDNs comply to the requirements of IEC 61000-4-6. Each CDN comes with individually measured data and a calibration certificate for the common mode impedance (fig. 1) and the voltage division factor (fig.3). Further typical data can be found in fig. 2 and fig. 4. 

A variety of calibration adapters and other accessories are available as option.

The capacitors at the AE-side are not equipped with discharging resistors

Remarks for the use of the supplied calibration adapters:
For the calibration it is mandatory to short all lines to create the common mode case. For this purpose the supplied calibration adapter (CA 6/4) is used. Due to the number of contacts, a relatively high insertion force is neccessary. Especially for unplugging a suitable lever with smooth surfaces is recommended. It is advisable to lift the adapter left and right in an alternating way several times to prevent unintentional bending.

The connection to ground can be accomplished using the ground plane of the CDN. Additionally there is a M4 thread located at the AE side to ground as well as a 4 mm socket to connect the device to ground.

To improve the operational safety the connections are carried out as security sockets. We recommend to use special 4 mm security plugs if you plan to design your own adapters. Those connectors can be purchased optionally.

Fig. 1: typ. EuT common mode impedance
Fig. 2: typ. phase EuT-Port, common- mode
Fig. 1: typ. EuT common mode impedanceFig. 2: typ. phase EuT-Port, common- mode
Fig. 3: typical voltage division factor RF-port to EuT port
Fig. 4: typ. isolation AE to RF-port
Fig. 3: typical voltage division factor RF-port to EuT portFig. 4: typ. isolation AE to RF-port

Schwarzbeck CDN AF6 Principal Circuitry
Fig. 5: Principal Circuitry CDN AF6

Specifications
Frequency range0.15 … 230 MHz
Common mode impedance150 kHz – 24 MHz: 150 Ω ±20 Ω
24 MHz – 80 MHz: 150 Ω +60 Ω / -45 Ω
80 MHz – 230 MHz: 150Ω ±60 Ω
Max. RF-test voltage (emf)30 V
RF-input-power6 W (continuous)
RF-input connector50 Ω BNC (female)
Voltage division factor RF-input – EuT-port9.5 dB 150 kHz – 80 MHz: ±1dB 80 MHz – 230 MHz: +3 dB / -2 dB
Input Voltage EuT (AE)160 V AC
250 V DC
Current rating4 A, DC
Housing materialAluminium
Housing dimensions216 x 105 x 108 mm
Weightca. 1400 g
EuT, AE connectors4 mm safety banana jacks
Included accessoriesCA 6/4
3 Pin shorting bar(2 units included by default)
Optional accessories
30 mm distance adapter
50 Ω to 150 Ω adapter

SR 30/4
SR 100-6W

Associated EMC Test Equipment