Loop Antennas

Quick Links

Information

Viewed products

Schwarzbeck FSH3D Isotropic 3-Axial RX Loop Antenna

New

  • 9 kHz - 200 (300) MHz
  • Light weight low attenuation radom
  • Outer diameter ca. 150 mm

More details

Request Quote/More Info

PDF Downloads

Test Equipment Description

The Schwarzbeck FSH3D consists of three identical, shielded loop antennas, which are aligned perpendicularly to each other. A shielded unit follows, containing the switching and amplifier circuitry, which is directly connected to the N-male output and fed by the Binder-connector with power supply and logic signals to select the desired loop-axis. The pin assignment is ready for operation with the popular handheld spectrum analyzers FSH3/6 and FSH 4/8 from Rohde & Schwarz.

A RF-transparent radome cover protects the FSH3D from dust and moisture. Because of its lightweight construction the FSH3D can be directly mounted to the spectrum analyzer. Alternatively the FSH3D can be held manually at its 50 mm shaft (sweeping method) or attached to a mast or tripod using the 22 mm tube (rotation method or raster method).

Electric Antenna FactorMagnetic Antenna Factor

Frequency [MHz]

Frequency [MHz]

Maximum Fieldstrength Measuring Range

Noise floor (RMS, BW: 200 Hz)

Frequency [MHz]

Frequency [MHz]

VSWR at N-Connector

Frequency [MHz]

Specifications
Frequency Range (nominal)9 kHz - 200 MHz
Frequency Range (useable)9 kHz - 300 MHz
Coaxial Connector (male)N
Power Supply5 V DC < 100 mA
Power Supply and Logic ConnectorBinder 99-0421-10-07 Serie 712
Antenna Factor for fictional E-fieldstrength42...70 dB/m
VSWRtyp. < 1.5 (20 kHz - 50 MHz)
Decoupling of the axistyp. > 20 dB
1 dB Compression at Outputtyp. 118 dBµV
Loop Diameter112 mm
Dimensions550 x 146 x 146 mm
Weight850 g
Environmental Conditions-10°C - +50°C, IP54

Control:

The active loop can be selected using the spectrum analyzer. Only one of three loops can be active at a time. The logic signals are labeled with TTL1 and TTL2, a logical 1 is represented by a voltage of 3.2 - 5 V. The alignment of the loop planes can be seen in the picture below. The loop of axis 2 lies in the cleavage plane of the radome. The loop planes 1 and 3 are standing vertically and perpendicular to each other, as depicted in the picture. The maximum field strength indication of a respective loop is achieved with the magnetic field lines going perpendicularly through the loop plane.

Binder 99-0421-10-07 Serie 712
Pin1 +5 V, max. 100 mA
Pin4 GND
Pin5 TTL1
Pin6 Connected to Pin 1
Pin7 TTL2

 TTL1TTL2
Rahmen / Loop 110
Rahmen / Loop 211
Rahmen / Loop 301
Reset00