Viewed products
5.7 inch color touch screen Full...
New
Standard
Warning: Last items in stock!
Availability date:
3ctest PFS Single Phase Power Fail Simulator simulates voltage dip, interruption and variation which are caused by power grid and transformer devices due to malfunction or sudden large change of load. It is required to test this kind of phenomena to check whether electronic and electrical device can sustain safety operating condition. PFS 2216 SV is equipped with two semiconductor switch controlled independently. It can withstand impulse current and each switch has electronic protection function.
General Data | ||
Display | 5.7 inch TFT touch screen | |
Power supply | 110V/220V, ±10%, 50/60Hz | |
Fuse | 6A | |
Max. Power consumption | 100W | |
Memory space | Infinite (PC) | |
Communication mode | Ethernet LAN, RJ45 | |
External control | D-sub 25p | |
External trigger input | BNC,5V TTL | |
CRO output | BNC,5V TTL | |
Input | BNC,5V TTL | |
Synchronization | 45-65Hz, 20-500V between dotted terminal & undotted terminal | |
Warning LED | 0-220VAC 10A/0-48VDC 3A Depending on external signal lamp | |
Safety circuit | Test only can be started when it is closed | |
Failure test | Display on LCD and interrupt the work at failure case | |
EUT voltage monitoring output | BNC, 100:1 coaxial output | |
EUT current monitoring output | BNC, 10A:1V coaxial output | |
Working indicator | LED, LCD front panel display | |
Grounding connection | Flat ground line | |
EUT power injection | 4mm banana line | |
Output mode | Standard 1m coaxial cable | |
Size | 19 inch/6U/4U or 22U cabinet | |
Weight(different weight for different model) | Eg. PFS 2210SVD / PFS 2210SD / PFS 2216SD weight: 40Kg; PFS 1130D weight: 24Kg or 100kg (with the power supply) | |
Temperature | 15–35°C | |
Humidity | 45%–75% | |
Air pressure | 86kPa–106kPa | |
Power fail test as per IEC6100-4-11&IEC61000-4-29,GB/T17626.29 for single phase AC and DC including voltage dips, short interruption and variation | ||
Power fail simulator | ||
EUT max. Voltage | PFS22xxSD, PFS22xxSVD series: 250V AC, 300V DC PFS05xxD series: 55VDC PFS11xxD series: 130VDC PFS12xxD series: 140VDC PFS22xxD series: 300VDC | |
EUT max. current | PFS2216SD, PFS2216SVD: 16A AC/DC continuous current; 20A continuous 5s; 40Acontinuous 3s; PFS2210SVD, PFS2210SVD: 10A AC/DC continuous current; 12A continuous 5s; 25Acontinuous 3s; PFSxx05D: 6A PFSxx010D:12A PFSxx20D: 22A PFSxx25D: 28A PFSxx30D: 33A PFSxx50D: 55A | |
EUT power supply | Single phase no more than 230V 50~60Hz±5% | DC no more than 300VDC Offer DC control |
EUT current and voltage measurement | LCD display BNC terminal: voltage 100:1, current 100A: 1V | |
Interrupt level | 0% | |
Dip level | 0%-100%(for accessories VVT/VMT series) 0%,40%,70%,80%(for accessories VVTxxxxSF series ) | |
Repetition rate | 0.3-9999 cycle or 5ms-9999ms | |
Duration | 50ms-50000ms | |
Test time | 1s-9999s | |
Rise and fall time | 1-5us(100Ω load) | |
Voltage variation level | 0-100% | |
Time for decreasing voltage | 500ms-60000ms(50%-100%)or 1-5us(100 ohm load) 1000ms-60000ms(0%-100%)or 1-5us(100 ohm load) | |
Time for increasing voltage | 500ms-60000ms(50%-100%) 1000ms-60000ms(0%-100%) | |
Time for reduced voltage | 10ms-99999ms | |
DC voltage level | 0-120% | |
DC voltage change continuous | 1ms-99999ms | |
DC voltage change interval | 50ms-50000ms | |
Synchronization | 0°- 360°,1° step or random | |
Trigger mode | Automatic, manual or external | |
Optional function | Additional module for voltage variation test and two DC mains | |
Peak Impulse Capacity of Simulator | ||
Rated current | Withstand current impulse capacity | |
PFS 2210SD | 500A | |
PFS 2210SVD | 500A | |
PFS 2216SD | 500A | |
PFS 2216SVD | 500A | |
PFS 2232SD | 500A | |
PFS 2232SVD | 500A | |
PFS 0505D | 24V/50A,48V/100A | |
PFS 0510D | 24V/50A,48V/100A | |
PFS 1110D | 24V/50A,48V/100A,110V/220A | |
PFS 1120D | 24V/50A,48V/100A,110V/220A | |
PFS 1125D | 24V/50A,48V/100A,110V/220A | |
PFS 1130D | 24V/50A,48V/100A,110V/220A | |
PFS 1150D | 24V/50A,48V/100A,110V/220A | |
PFS 2210D | 24V/50A,48V/100A,110V/220A | |
PFS 2225D | 24V/50A,48V/100A,110V/220A | |
PFS 2250D | 24V/50A,48V/100A,110V/220A | |
Simulator with rated current no more than 16A meets the load capacity specified in IEC 61000-4-11 | ||
Rated current | Load current capacity | |
PFS 2210SD PFS 2210SVD | 12A AC/DC continuous current; 15A continuous 5s; 25Acontinuous 3s; | |
PFS 2216SD PFS 2216SVD | 18A AC/DC continuous current; 20A continuous 5s; 40Acontinuous 3s; |