RF Field Strength Probes & Meters

Quick Links

Information

Viewed products

EFP40G Isotropic Electric Field Probe 1 MHz - 40 GHz View larger

EFP40G Isotropic Electric Field Probe 1 MHz - 40 GHz

New

  • 1 MHz - 40 GHz frequency range
  • 2-1000 V/m 
  • Operates with RFM high performance broadband radiation field meter
  • Fast sampling detection for MIL-STD-461 & most EMC, RF, communication safety standards

More details

Typically In Stock

Request Quote/More Info

PDF Downloads

EFP40G Datasheet

Rent or Buy the EFP40G Isotropic Electric Field Probe 1 MHz - 40 GHz at The EMC Shop Today!

PDF Downloads (354.86k)

Specifications

Frequency range1 MHz-40 GHz
Dimensions (LxD): 12.0"x2.25" (310x58 mm)
Weight0.5 lb, 230g.

Test Equipment Description

EFP40G Isotropic Electric Field Probe 1 MHz - 40 GHz Details

The EFP40G Isotropic Electric Field Probe operates in 1 MHz to 40 GHz covering most test requirements EMC and RF safety standards for industrial, military and radar communication applications. High overload field strength limit and operation up to 40 GHz makes this probe a versatile measuring tool for all applications. This including the microwave K-Band which requires field stregeth of 18 to 26.5-GHz.     

  • Diode based: 3-Axial isotropic sensor.

  • Sensitivity: 1.2 V/m.

  • Linearity: +/- 0.5 dB (2-400 V/m),

    • +/- 1 dB (400-1000 V/m).

  • Dynamic range: >56 dB (single range).

  • Frequency response (without correction factors):

    • +/- 3 dB (1 MHz–3 GHz), -3 dB @ 1 MHz

    • +6/-0 dB (3 GHz–18 GHz)

    • +/- 3 dB (18 GHz–40 GHz)
  • Frequency response (with correction factors):
    • +/- 1 dB (1 MHz–12 GHz),
    • +/- 2 dB (12 GHz–40 GHz).
  • Rotational isotropicity:
    • +/- 0.5 dB (1 MHz-3 GHz),
    • +/- 1.5 dB (3 GHz – 12 GHz)
    • +/- 3 dB (12 GHz – 40 GHz)
  • Calibration accuracy (at specified freq.) <1 dB.
  • Temperature: 5-40 °C,
  • Humidity: RH 10-90%, non-condensing.
  • Temperature error: <0.05 dB/°C (field > 5 V/m).
  • UK NPL traceable calibration.
  • Small size: LxD: 9.0x2.25 inch, 230x58 mm.
  • Weight: 0.2 lb, 100 g.
  • Operates with RFP-05 meters