Viewed products
300 kHz to 26.5 GHz Up to 66 dB...
New
Typically Stocked
Warning: Last items in stock!
Availability date:
Frequency Range | 300 kHz to 26.5 GHz |
Standards Met | IEC/EN 61000-4-3 |
Application/Test Type | Monitoring electric field strengths in GTEM's, Anechoic Chambers and other RF rooms for radiated immunity testing. |
RF Field Probes E-Field | |
Parameter | Value |
Sensors | Six monopoles |
Measurement Mode | X-Y-Z axis simultaneous sampling |
Sampling Rate | 22 to 0.03 Samples/s depending on filter settings |
Digital Filtering | Low-pass 2.3 to 28 Hz; 80 dB notch 50/60 Hz |
Battery | 3V, 5 mA/h, Li-Mn rechargeable |
Operation Time | 80 hours @ 0.4 S/s, 28 Hz filter; 60 hours @ 5 S/s, 28Hz filter |
Recharging Time | 48 hours for full operation time Partial charging allowed (8 hrs recharging allows 12 hrs operation @ 5 S/s 1 hr recharging allows 5 hrs operation @ 0.4 S/s |
Flatness | Frequency correction OFF: 10 to 18000 MHz: 1.8 dB; 3 to 23000 MHz: 3.2 dB; Frequency correction ON: typically 0.3 to 26500 MHz: 0.4 dB |
Dynamic Range | 0.4 to 800 V/m; 66 dB single range |
Overload | 1,600 V/m |
Internal Data Memory | Serial nr. - Calibration factors & date Firmware version |
Communication | Bidirectional fiber optic link |
PC Software | Displays field, temperature, battery Sampling, filters, averaging, frequency setting Manual data logging as text file |
Probe Holder Tip | 1/4 20 UNC female |
Battery Voltage Reading Resolution | 10 mV |
Temperature Reading Resolution | 0.1 °C |
Resolution | 0.01 V/m |
Dimensions | 45mm (1.7 inches) |
Weight | 22 grams (0.77 ounces) |
The all-in-one that sets the standard for miniature optically coupled broadband E-field isotropic probes. Exceptionally small-sized spherical symmetrical configuration, lightweight and miniaturized electronics combined with excellent RF characteristics make the Narda PMM EP-604 E-field Probe 800 V/m CW to 26.5 GHz is the ideal solution for all EMC/EMI applications (chambers and TEM/GTEM cells), biology and materials research and wherever fast and accurate measurements with negligible or minimum interference of the probe to the field under measure are essential. Accredited calibration on request.