Schloder SFT 415 - CS Calibration Set for CCC View larger

Schloder SFT 415 - CS Calibration Set for CCC

New

  • For Capacitive Coupling Clamp Burst
  • IEC 61000-4-4, 2012
  • Transducer Plate with Connection Adapter
  • Optionally 50 ohm Attenuator

More details

Request Quote/More Info

PDF Downloads

Test Equipment Description

Calibration - Set

The capacitive coupling clamp is specified in the standard IEC / EN 61000-4-4, chapter 6.4. It provides the ability of coupling the fast transients / bursts to the circuit under test without any galvanic connection to the terminals of the EUT´s ports, shielding of the cables or any other part of the EUT.

With the release of the standard IEC 61000-4-4, 2012 the capacitive coupling clamp is also to verify at their characteristics and to calibrate.

Calibration of the capacitive coupling clamp

Required accessories:

  • Calibration-Set (SFT 415-CS)
    Transducer plate with connection adapter Support for the measurement attenuator
  • Measurement attenuator 50 Ohm (SFT 450-1)

Measurement equipment that is specified as suitable to perform the calibrations defined in IEC 61000-4-4, chapter 6.2.3 (Burst generator verification) shall also be used for the calibration of the characteristics of the capacitive coupling clamp.

Technical Data
Transducer Plate with Connection Adapter
Materialstainless steel plate, conducting
Dimension120 mm x 1050 mm
Material thickness0,5 mm
Isolationdielectric foil with thickness of 0,5 mm, all around
Insulation voltagemin. 2,5 kV – on all sides
Connection for 50 ohmFischer HV-female connector
Other connectors (BNC HV) on request
Total length Set1085 mm (with adapter)
Weight720 g
Working temperature0 - 40° Celsius
Relative humidity0 - 60%