Viewed products
Construction as per IEC 61000-4-4 See...
New
Warning: Last items in stock!
Availability date:
Voltage Rating | up to 5.0 kV |
Standards Met | IEC / EN 61000-4-4 |
RF Connector | Fischer HV Connector |
Application/Test Type | Electrical Fast Transients on Data Lines (EFT Burst) |
Dimensions | 100 cm x 14 cm x 14 cm |
Weight | 7.5 kg |
EFT Coupling Clamps as per IEC 61000-4-4 | |
Parameter | Value |
Cable Diameter | up to 40 mm (1.57 in) |
Coupling Length | 1 m (39.4 in) |
The Schloder capacitive coupling clamp is specified in the standard IEC / EN 61000-4-4, chapter 6.4. It provides the ability of coupling the fast transients / bursts to the circuit under test without any galvanic connection to the terminals of the EUT´s ports, shielding of the cables or any other part of the EUT.
Transducer plate with connection adapter Support for the measurement attenuator
Measurement attenuator (optional)
Place the calibration set into the capacitive coupling clamp. Connect the 50 ohm measurement attenuator with the transducer plate and the other end with the oscilloscope. On the opposite site of the coupling clamp must connected the burst generator.
The calibration is performed with the generator output voltage set to 2000 V.
The (Burst) waveform characteristics shall meet the following requirements: