EMC Test Equipment

Quick Links

Information

Viewed products

Schloder SFT 1400 EFT/Burst Generator up to 5 kv, IEC/EN 61000-4-4 View larger

Schloder SFT 1400 E125 kHz Burst Generator

New

  • Frequency up to 125 kHz 
  • Changing of all Parameters during a Burst Test 
  • IEC 61000-4-4 up to 5 kV
  • Single Spike to Continuous Burst
  • 5000 Pulses/Sec – 500/Package 
  • Time Sccurate Triggering 
  • Various Special Function

More details

Typically In Stock

Request Quote/More Info

PDF Downloads

Specifications

Burst (EFT) 5/50 ns
Pulse conforms to IEC/EN 61000-4-4
ParameterValue
Pulse amplitude100 V - 5000 V
Burst frequencysingle up to 125 kHz
Polaritypositive / negative / alternate
Test duration10 - 1000 ms
CRO TriggerRS 232 Remote control
Repetition rate0.01 - 100 ms
CouplingIntegrated in the test generator, coupling of the noise pulses to the EUT’s power mains
Rise time5 ns ±30%
Burst time10 - 1000 ms
Dimensions19″ housing. 3 HE
Weight9 kg

Test Equipment Description

Schloder SFT 1400 E125 kHz Burst Generator

The test generator simulates quick transient noise interference as they are defined in the standards IEC 61000-4-4 and EN 61000-4-4. The single pulses show a very short rise-time (5ns) and due to this a wide RFspectrum up to 300 MHz. RF-interferences are the result.

Special function
The generator includes several special functions such as "Real Burst" which simulates the natural appearance of the burst phenomena or "Sweep" to simulate the bouncing of an electrical contact. The functions "IFM" and "DFM" (increasing and decreasing frequency) are powerful instruments to investigate resonance or saturation effects in the tested device.

Easy operation
The clearly arranged front panel with the generator settings allows a time-saving and optimised testing. The standard test level 1, 2, 3 and 4 are stored in the memory function on position 1 – 4. Additional custom made setups can be stored in the memory function.

Associated EMC Test Equipment

30 other products in the same category: