Standards Met | IEC 6100-4-5, ANSI/IEEE C62.41 Cat. B, UL 1449 (8/98) , IEC 61000-4-4 |
Pulse voltage (open circuit) | 1.2/50µs, 200V - 6.6kV -5 +10% in 1 volt steps |
Combination wave pulse 1, 2/50 - 8/20 μs (Hybrid-Surge pulse) Pulse conforms to IEC/EN 61000-4-5 | |
Parameter | Value |
Pulse current (short circuit) | 8/20µs, 100A - 3.3kA -0 +10% with 2 ohm effective source impedance. |
Coupling | 1 PH / 15 (16) Amps |
Repetition time | Adjustable from 300ms to 5s; 1ms resolution |
Pulse repetition | 1 shot/12 seconds; Rise Time Tolerance ±30% for voltage; ±20% for current |
Phase synchronization | ±15° with optional coupler/decoupler |
Coupling | Model E4551 - Single Phase 15 Amps |
Test duration | ±20% Tolerance |
Test duration | 1.0ms to 20ms; 1.0ms resolution or 1 to 200 pulses |
Burst (EFT) 5/50 ns Pulse conforms to IEC/EN 61000-4-4 | |
Parameter | Value |
Pulse amplitude | 200V to 4.4kV (5V Steps |
Burst frequency | Adjustable from 1kHz to 1MHz up to 4.4kV |
Polarity | Positive, negative, alternating |
Burst time | 1 to 360 sec.; 1 sec. resolution 1 to 240 min.; 1 min. resolution 1 to 24 hours; 1 hour resolution |
The Thermo Fisher Keytek ECAT Simulator systems are a sesoned favorite being used for electromagnetic Surge and EFT testing for more than 2 decades. The controller has an easy-to-use Windows based application software for quick implementation of test routines and standards, with automoatic report generation. A multi-level system interlock architecture provides maximum safety to keep first time users safe.
This rental ECAT system is suitable for users looking to evaluate DUT in a precompliance environment for passing Surge and Burst (Electrical Fast Transient) immunity test standards such as IEC 61000-4-4, IEC 61000-4-5 and ANSI C62.41. The single phase, 15 amp coupling network is suitable for most commercial electronics and lower powered electronics.