Viewed products
Rent EM Test NX5 here. Model:...
New
In Stock
Warning: Last items in stock!
Availability date:
Pulse voltage (open circuit) | 160 V - 5,000 V ± 10% |
Combination wave pulse 1, 2/50 - 8/20 μs (Hybrid-Surge pulse) Pulse conforms to IEC/EN 61000-4-5 | |
Parameter | Value |
Pulse current (short circuit) | Max 2,500 A ± 10% |
Coupling | external / internal |
Repetition time | 10 ms - 9,999 ms |
Pulse repetition | max. 1 Hz (1 s - 9999 s) |
Phase synchronization | 0° - 360°, resolution 1° (16 - 500 Hz) |
Polarity | positive / negative / alternate |
Coupling | external / internal |
Phase synchronization | 0° - 360°, resolution 1º |
Dips, Interrupts & Variations Conforms to IEC/EN 61000-4-11, IEC/EN 61000-4-29 | |
Parameter | Value |
Test duration | 1 - 99,999 or endless, selectable |
Test duration | 0:01 min - 99:59 min or endless |
Peak inrush current capability | greater than 500 Amps |
Burst (EFT) 5/50 ns Pulse conforms to IEC/EN 61000-4-4 | |
Parameter | Value |
Pulse amplitude | 200 V to 5.5 kV 100 V to 2.75 kV (50 Ω matching system) |
Burst frequency | 0.10 ms - 9999 ms |
Polarity | positive / negative / alternate |
Switching times | greater than 1 us less than 5 us into a 100 ohm resistive load (SVP 100) |
Event time | 20 us - 9,999 s |
Test duration | 20 us - 9,999 s |
Repetition time | 10 ms - 9,999 s |
Phase synchronization | 0° - 360°, resolution 1° (16 - 500 Hz) |
Burst time | 0.1 kHz - 1,000 kHz |
The EM Test NX5 is a versatile tester to address transient and power fail requirements for both international and commercial standards. Featuring an easy-to-use color touch screen, the NX5 provides an economical solution for pre-compliance immunity testing as well as full-compliance testing and CE Marking.
The EMC Shop supplies a large range of accessories for various applications such as magnetic field tests and more.
The standard configuration available when you rent from The EMC Shop includes capabilities of EFT/Burst (IEC 61000-4-4) up to 5.5 kV, Surge (IEC 61000-4-5) up to 5.0 kV and Powerfail (IEC 61000-4-11), with a built in single phase coupler rated to 300 Volts (p-n) AC + DC and 16 Amps.
The EM Test NX5 can be operated manually from the intuitive front touch screen or remotely via its built-in Ethernet, USB or optical interface.
The EM Test NX5 has clearly structured windows and operation features, comprehensive standards library along with the flexibility to easily generate user specific test sequences are the main features of iec.control software.
Extensive reporting capabilities help the user to create test reports that meet international requirements. The iec.control is supported by Windows 7, Windows 8 and Windows 10.
Remote control is achieved either via Ethernet or optical interface with USB connector at the PC side.