Viewed products
10 kHz–400 MHz Frequency Range Bulk...
New
Complete testing solutions to the following standards:
Warning: Last items in stock!
Availability date:
Amplifier Research CI00400AM4 M1 through M7 RF Conducted Immunity System 100 Watts nominal 10 kHz–400 MHz
The Amplifier Research CI00400A is a fully self-contained state of the art system designed to test RF Conducted Immunity.
The CI00400AM4 has an additional power head for testing automotive standards such as Ford, GM, Chrysler, BMW, Renault, and ISO11452-4.
The system contains a signal generator, 3 channel power meter, 100W nominal AR amplifier 10kHz to 400MHz, directional coupler, and control software. Everything is contained in a single housing, which eliminates setup issues. This system provides the versatility needed for every test laboratory and equipment manufacturer. The RF amplifier and the signal generator can be used independently of the system. If special needs arise or standards were to change a larger amplifier can be connected to the system. The use of spectrum analyzers and monitoring equipment may also be controlled by the software.
EXPORT CONTROLLED DATA
These commodities, technology, or software are controlled for export in accordance with the U.S. Export Administration Regulations. Diversion contrary to U.S. law is prohibited.
RF Conducted Immunity Generator with a 100 watt power amplifier with a frequency range of 10 kHz - 400 MHz. Includes a three-channel power meter with three channels used for forward power, reverse power and signal level monitoring, signal generator (9 kHz - 1.2 GHz), directional coupler, additional power head and control software with custom test editor. Meets MIL-STD 461, DO160 and ISO 11452-4 standards.
Signal Generator Specifications | |
Frequency range resolution | 9 kHz to 1.2 GHz 1Hz |
Power range resolution | -140 to +13 dBm 0.1dB |
Modulation | AM, FSK. FM, Phase, External Pulse |
RF Amplifier Specifications | |
Frequency range | 10 kHz to 400 MHz |
Power rating | 100 Watts Nominal |
1dB compression | 75 Watts Nominal |
Harmonic Distortion | -20dBc at 50 Watts |
Mismatch tolerance | 100% of rated power without fold back. Will operate without damage or oscillation with any magnitude of source and load impedance. |
Gain | 51dB minimum |
Connections | |
RF Out | Type N (front) |
Monitor Port In | Type N (front) |
Signal Generator Out | Type N (rear) |
Directional Coupler In | Type N (rear) |
Amplifier Out/In | Type N (rear) |
Pulse In | BNC (rear) |
Communication | GPIB (IEEE 488) (rear) |
Directional Coupler Fwd Out | Type SMA (rear) |
Directional Coupler Fwd In | Type SMA (rear) |
Directional Coupler Rev Out (with Option 1) | Type SMA (rear) |
Directional Coupler Rev In (with Option 1) | Type SMA (rear) |
Monitor Port Out | Type SMA (rear) |
Monitor Port In | Type SMA (rear) |
Power Meter Calibration Port Out | Type SMA (rear) |
General | |
Power | 115/230 VAC 50/60 Hz, single phase 16A |
Breaker | 2 pole, 20A |
Cooling | Active cooling, air ventilation |
Environmental conditions | 10°C - 40°C |
Dimensions | 50.3 x 42.2 x 52.1 cm 19.8 x 16.6 x 21.7 in |
Weight | 22.7 kg (50.0 lb) |
Weight CI00400AM5 | 17.3 kg (38.2 lb) |
Control Software | emcware® Software Suite |
PC Requirements | |
Computer | 2GHz Single Core Processor |
Operating system | Windows XP, Windows Vista, Windows 7 |
RAM | 1 GB Minimum |
Screen Resolution | 1024 x 768 |
Ports | 2 available USB 2.0 ports |
Options | |
1 | Additional power head to add the ability to monitor reverse power |
2 | Data acquisition card. |
3 | Laptop PC with software preinstalled |
4 | Amplifier removed; requires use of external amplifier |